Waters Acquiry TUV Detector

 4.555,00 excl. tax

1 in stock

SKU: RBN 22759 Category:
Description

Waters Acquiry 186015022 TUV detector HPLC UPLC

Wavelength range: 190 to 700 nm

Bandwidth: <5nm

Wavelength Accuracy: ±1.0nm

Wavelength repeatability: ±0.1nm

Single λ dry noise: <6 μAU (at 230 nm, 1 second digital filter, 30 second segments, 2 Hz data rate, medium shunt in place)

Two-channel noise (dry): 10.0 x 10-5 AU (measured at 230 nm, operating at 230 to 280 nm with 2 second filter at 2 Hz data rate; dry, medium shunt in place)

Linearity: <5% at 2.5 AU, propylparaben at 257 nm

Drift: 5.0 x 10-4 AU/hr. max (ΔΤ = +2 °C/h) at 230 nm, after 1 h. warm-up, with medium shunt

Sensitivity setting range: 0.0001 to 4.0000 AUFS

Filter setting range: single wavelength; 0.1 to 5.0 seconds, dual wavelength; 1 to 50 seconds

Lamp source: 30 W High Brightness Deuterium lamp, 0.5 nm aperture, pre-aligned, accessible from the front

Photodiodes: 2 silicon photodiodes (matched pair)

Second order filter: automatic for wavelengths > 370nm

Wavelength Calibration Filter: Erbium filter, used at startup or on demand

Nitrogen purge: purge fitting present on optics table

Path length: 10mm

Cell volume: 0.5 µL

Pressure limit: 69 bar