Wavelength range: 190 to 700 nm
Bandwidth: <5 nm
Wavelength accuracy: ±1.0 nm
Wavelength repeatability: ±0.1 nm
Single λ dry noise: <6 μAU (at 230 nm, 1-second digital filter, 30-second segments, 2-Hz data rate, medium shunt in place)
Dual channel noise (dry): 10.0 x 10-5 AU (measured at 230 nm, running at 230 to 280 nm with 2-second filter at 2-Hz data rate; dry, medium shunt in place)
Linearity: <5% at 2.5 AU, propylparaben at 257 nm
Drift: 5.0 x 10-4 AU/hr. max (ΔΤ = +2 °C/hr.) at 230 nm, after 1 hr. warmup, using medium shunt 
Sensitivity setting range: 0.0001 to 4.0000 AUFS
Filter setting range: Single-wavelength; 0.1 to 5.0 seconds, Dual-wavelength; 1 to 50 seconds
Lamp source: 30-W High Brightness Deuterium lamp, 0.5 nm aperture, pre-aligned, front accessible
Photodiodes: 2 silicon photodiodes (matched pair)
Second-order filter: Automatic for wavelengths > 370nm
Wavelength calibration filter: Erbium filter, used at startup or on demand
Nitrogen purge: Purge fitting present on optics bench
Path length: 10 mm
Cell volume: 0.5 μL
Pressure limit: 69 bar