BD FACSCanto II Flow Cytometer
The BD FACSCanto II from BD Biosciences is an advanced, compact flow cytometer designed for multiparametric cell analysis. The system is intended for clinical laboratories, research institutions, and biomedical applications where accurate, reproducible, and reliable analytical results are essential.
The FACSCanto II utilizes a two- or three-laser configuration and is capable of detecting up to 8 fluorescent colors. This allows for the simultaneous analysis of multiple cellular characteristics, making the instrument particularly well-suited for immunophenotyping, cell characterization, and hematological research.
A key feature is its fixed optical alignment (Fixed Alignment Flow Cell), which eliminates the need for manual laser alignment by the user. This reduces start-up time and contributes to consistent, reproducible measurement results.
The advanced optical system optimizes signal detection, providing high sensitivity and resolution for multicolor analyses. With an automated fluidics system, comprehensive quality control, and high processing speed, the BD FACSCanto II is well-suited for laboratories with high sample throughput.
Key Features
Flow cytometric analysis with up to 8 fluorescent colors.
2- or 3-laser configuration.
Three available laser wavelengths: 405, 488, and 633 nm.
Fixed optical alignment; no manual adjustment required.
8 fluorescence detectors (PMTs).
High sensitivity and resolution for multicolor analyses.
Sample processing rate of up to 10,000 events per second.
Automated fluidics system.
Suitable for clinical diagnostics and scientific research. Reproducible measurement results and comprehensive quality control. Technical specifications
1. Laser system
Specification Value
Manufacturer BD Biosciences
Model FACSCanto II
Device type Flow cytometer
Laser configuration 2 or 3 lasers
Violet laser 405 nm solid state diode
Violet laser power 30 mW (fiber output)
Blue laser 488 nm solid state
Blue laser power 20 mW
Red laser 633 nm HeNe
Red laser power 17 mW
Laser configuration Spatially separated laser beams
Laser spot size 9 × 65 µm (elliptical)
Optical alignment Fixed (Fixed Alignment)
2. Flow cell and optical system
Specification Value
Flow cell Rectangular quartz flow cell
Flow cell dimensions 180 × 430 µm
Collection optics Optical-gel coupled lens
Numerical aperture 1.2 NA
FSC resolution 1.0 µm
SSC resolution 0.5 µm
Fluorescence detection Reflective optics
Fluorescence detectors 8 PMTs
Detector configuration 4-2-2
3. Fluorescence detection
Laser Fluorochromes
Violet (405 nm) Pacific Blue, AmCyan
Blue (488 nm) FITC, PE, PerCP, PerCP-Cy5.5, PE-Cy7
Red (633 nm) APC, APC-Cy7
Detector bands:
Laser Detector Band
Violet 450/50nm
Violet 502–525 nm
Blue 530/30nm
Blue 585/42nm
Blue >670 nm
Blue 780/60nm
Red 660/20nm
Red 780/60nm
4. Detection and sample capacity
Specification Value
FSC detector Photodiode
SSC detector PMT
Fluorescence detectors 8 PMTs
FITC sensitivity <100 MESF
PE sensitivity <50 MESF
Maximum particle size 50 µm
Sample flow rate 10–120 µL/min
Maximum acquisition rate 10,000 events/sec
Fluorescence parameters 6 compensated parameters
Scatter parameters 2
Sample dead volume 30 µL
Compatible sample tubes BD Falcon 12 × 75 mm
5. Dimensions and weight
Cytometer – Fluidics cart
Depth 610mm – 610mm
Width 910mm – 790mm
Height 640mm – 640mm
Weight 145 kg – 51 kg
Total system configuration:
Cytometer: 145 kg
Fluidics cart: 51 kg
Total weight: 196 kg