Max. number of plates: Up to 7 microtiter plates at RT
Barcode: Barcode identification of samples, reagents and microtiter plates
Barcode laser: Scanner for racksystem
Max. number of samples: Max. 100 tubes, up to max. 16 mm outer diameter, length up to 100 mm
Reagents: SERION racks
Dilution positions: Up to 4 dilution plates
Max. number of tip racks: Up to 5 tip racks with 300 μl or 1100 μl disposable tips
Pipetting system: Liquid pipettor for disposable tips
Volumes min./max.: For 300 μl tips 10–300 μl, for 1100 μl tips 301–1000 μl
Accuracy: < 15 % CV for 25 μl volumes; < 5 % CV for 100 μl volumes
Precision: < 5 % CV for 25 μl volumes; < 2.5 % CV for 100 μl volumes
Incubation capacity: 4 independent chambers with agitation function
Incubation range: Between RT +7 °C up to max. 50 °C with temperature control
Temperature: +/–2 °C
Capacity of buffers: Max. 4 positions
Dispense volume: 200–999 μl/well
Residual volume: < 2.5 μl in U-shaped bottom; < 4 μl in flat bottom
Precision: 5 % CV for 300 μl volumes
Fluid alarms: Automated warning in case of lack of reagents or filled waste container
Spectral range: 400–700 nm
Dynamic range: 0.0 OD up to 3.0 OD
Linearity: 0-2.000 OD +/- 1%
Accuracy: +/– 0.005 OD or 2.5 %
Reading time / 96 wells: < 15 sec.
Number of filters: Up to 8 filter positions (Standard 405, 450, 620 nm)
Dimensions (WxDxH): 1140 x 750 x 1000 mm
Weight: 130 kg
Software included
| Weight | 0,0 kg |
|---|---|
| Brand | Other brands |
| Warranty | 6 months |
| Condition | Used, in good condition |
| Date of Manufacture | 2015 |