DiaSorin ETI-Max 3000 Fully Automated Microplate Analyzer

 2.500,00 excl. tax

1 in stock

SKU: LM 24971 Category:
Description

The ETI-MAX 3000 is a fully automated microtiter plate analyzer performing the complete sample processing (sample pre-dilutions, sample and reagent dispensing, incubations, wash processes, plate transports) as well as the photometric measurement and evaluation.

The instrument is controlled by Windows PC software. This software allows the user to process the pre-defined assays of DiaSorin as well as assays programmed by the user. The clear structure with intuitive user-guidance allows simple and quick operation of daily routine jobs as well as programming of user-specific assays.

Liquid handling: 1 ml thinner

Disposable tips: 300 or 1100 µl

Pipette range: Simultaneous operation of 4 plates

Precision (samples and reagents): CV < 8.0% at 10 µl, CV < 2.5% at 100 µl

Level sensor: Electronic

Clot detection: Yes

Mixing: Yes (pre-dilution tube and microplate)

Multipipetting: Yes (sample, control and reagent)

Pipetting time samples: < 18 min/96 wells (100 µl/well)

Pipetting time reagents: < 4 min/96 wells (100 µl/well)

Carry-over: None

Identification: Barcode scanner for primary vessels, controls and reagents, Barcode scanner for microtiter plates

Tube: Diameter 10-16 mm, Height 55-100 mm

Barcode type: 2/5 interleaved, UPC A & E, 2/5 IATA, 2/5 Industrial, EAN 8 of 13, Code 128, EAN 128, Pharmacode, EAN addendum, 2 of 5, Code-a-bar

Capacity: Maximum 240 sample tubes

Capacity: 4 separate incubators

Temperature range: 5°C above room temperature to 50°C

Accuracy: ± 1°C average of the plate

Consistency: ± 0.7°C/whole plate

Shaking: Lengthwise

Capacity: Up to 4 wash buffers

Wash head: 1 x 8

Pipette volume: 200 – 2500 µl/well, depending on the assay

Precision: ± 5% CV at 300 µl

Residual volume: < 2.5 µl in round-bottom wells, < 4 µl for flat-bottom wells

Level sensor buffer: Yes

Level sensor waste container: Yes

Wash programs: 1 to 9, depending on the assay

Incubation time: 1 to 999 seconds, depending on the assay

Pipette pressure: Adjustable for each assay individually

Measurement: Vertical with photodiodes, Extinction or kinetic measurement

Channels: 8

Operation: Single or double, with second measurement when the measuring range is exceeded

Spectrum: 400 – 700 nm

Filter: Up to 8 positions are available, of which 5 are already on board (405, 450, 492, 550, 620 nm)

Measurement time: < 10 seconds

Dynamic range: – 0.100 to 3,000 absorption units

Linearity: 0 –2,000 absorption units ± 1.0%

Precision: ± 0.005 absorption units or 2.5%

Dimensions (WxDxH): 1130 x 760 x 1000 mm

Weight: 130 kg

Specifications
Weight 0,0 kg
Warranty

6 months

Condition

Used, in good condition