Very well maintained, near mint condition, fully operational and tested
Single Cell Holder for 1 cm path length cuvettes.
Windows 10 and 11 compatible software.
Onsite installation, configuration and software training by highly skilled technicians (for buyers within the Benelux only)
Step by step installation guide and online support for buyers outside the Benelux
Real double beam technology
Source: Deuterium and Halogen (tungsten)
Wavelength range: 190 – 1000 nm
Absorption range: +/- 4 A
Stray light at 220 nm (Nal): < 0.02 %T
Stray light at 340 nm (NaNO2): < 0.02 %T
Stray light at 198 nm (KCl): < 1 %T
Bandwidth: 0.5, 1, 2 , 5 and 20
Resolution test (0.02 % Toluene in hexane, ref. EP): > 1.6
Wavelength accuracy (Holmium oxide solution, ref. NIST 2034): +/- 0.3 nm
Wavelength Accuracy (656.1nm): +/- 0.1nm
Wavelength reproducibility (656.1 nm, 10 measurements): <0.01 nm
Photometric accuracy (1 A, ref. NIST 930e): +/- 0.004 A
Photometric reproducibility (1 A): < 0.001 A
Photometric drift: < 0.0003 A/h at 700 nm
Photometric noise (RMS): < 0.00005 at 700 nm
Scan speed: 1 – 3000nm/min
Beam height: 15 mm
USB interface
Complies with European Pharmacopoeia
Dimensions (WxDxH): 491 x 538 x 258 mm
Weight: 23.5 kg
Weight | 0,0 kg |
---|---|
Warranty | 0 months |
Condition | Used, in good condition |
Brand | Perkin Elmer |