The appliance is intended for the creation and maintenance of constant environmental conditions for application in the field of in vitro fertilisation (IVF), especially for the incubation of oocytes, spermatozoa and zygotes in special culture dishes for IVF application as well as for gene expression and the biosynthesis of RNA and proteins.
Working-temperature range: 5°C above ambient temperature up to 50°C
Standard sterilisation program: 60 minutes at 180°C (without removing the sensors)
Setting temperature range: 18 to 50°C
Setting accuracy temperature: 0.1°C
Temperature sensor: 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error
Temperature variation in chamber (at 37°C): +/- 0.3 K
Temperature fluctuations with time (at 37°C): +/- 0.1 K
Humidity control: Peltier element limits the value of relative humidity in the interior to 93 % rh +/- 2.5 %
Setting accuracy humidity: 0.5 % rh
CO₂ control: Digital electronic with dual beam NDIR system, with auto-diagnostic system and acoustic fault indication, barometric pressure compensation
Adjustment range CO₂: 0 to 20%
Setting accuracy CO₂: 0,1%
Variation in time CO₂: +/- 0.2%
Max. number of internals: 10 (2 included)
Max. loading of chamber: 120 kg
Max. loading per internal: 15 kg
Volume: 156 l
Internal dimensions (WxHxD): 560 x 700 x 400 mm
Dimensions (WxHxD): 719 x 1070 x 591 mm
Weight: 90 kg
Weight | 0,0 kg |
---|---|
Warranty | 6 months |
Condition | Used, in good condition |
Brand | Memmert |