Included sensors:
Oxygen/temperature sensor – PODOA-PODOB
Measurement principle: Luminescence-based optical measurement
Measurement ranges: 0.00 to 20.00 mg/L or ppt, 0-200% (saturation)
Resolution: 0.01
Accuracy: +/- 0.1 mg/L or +/- 1 %
Response time: 90% of the value in less than 60 seconds
Water movement: No circulation required
Temperature compensation: Via an NTC thermistor
Temperature measurement range: 0 – 60°C
Temperature resolution: 0.01
Temperature accuracy: +/- 0.5 °C
Storage temperature: – 10°C to + 60°C
Signal interface Modbus: RS-485 (SDI-12)
Measurement refresh rate: < 1 second maximum
pH/Redox/temperature sensor – PPHRA-PPHRB
Measurement principle (pH): pH/reference combined electrode: special glass, Ag/AgCI reference. Gel (KCl) electrolyte
Measurement range: 0 – 14 pH
Resolution: 0.01 pH
Accuracy: +/- 0.1 pH
Measurement principle (Redox): Redox/reference combined electrode: Platinum disk, Ag/AgCI reference. Gel (KCl) electrolyte
Measurement range: – 1000,0 à + 1000,0 mV
Resolution: 0.1 mV
Accuracy: ± 2 mV
Measurement principle (T°C): NTC thermistor
Operating temperature: 0.00 °C to + 60.00°C
Resolution: 0.01 °C
Accuracy: ± 0.5 °C
Storage temperature: 0°C to + 60°C
Ingress rating: IP 68
Signal interface: Modbus RS-485 as standard (SDI-12 as an option)
Measurement refresh rate: < 1 second maximum
Dimensions (HxWxD): 43 x 122 x 205 mm
Weight: 560 g
| Weight | 0,0 kg |
|---|---|
| Warranty | 1 month |
| Condition | Demo |