Working-temperature range: Ambient +5°C up to +50 °C
Setting temperature range: +18°C to +50°C
Setting accuracy temperature: 0.1°C
Temperature variation in chamber (37°C): ±0.3 K
Temperature fluctuations with time (37°C): ±0.1 K
Standard sterilisation program: 60 minutes at 180°C (without removing the sensors)
Temperature sensor: 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error
Humidity control: Humidity limitation thanks to a Peltier element; when water dish is full and inserted, the Peltier element limits the value of relative humidity in the interior to 93% rh ± 2.5 %
Setting accuracy humidity: 0.5 % rh
CO2 control: Digital electronic CO2 control with dual beam NDIR system, with auto-diagnostic system and acoustic fault indication, barometric pressure compensation
Adjustment range CO2: 0 to 20 %
Setting accuracy CO2: 0,1%
Variation in time CO2: ±0.2%
Door: inner glass door with opening (8mm Ø) to take gas sample
Capacity: 56 l
Internal dimensions (WxHxD): 400 x 425 x 330 mm
Dimensions (WxHxD): 559 x 795 x 521 mm
Weight: 55kg
Including software
Weight | 0,0 kg |
---|---|
Warranty | 0 maanden |
Condition | Zo goed als nieuw |
Brand | Memmert |