Dionex UltiMate 3000 VWD-3400 Variable Wavelength Detector

 1.140,00 excl. tax

Not available

SKU: RBN 14517 Category:

Noise, single wavelength: < ±2.0 μAU (typical) at 230 nm, time constant 2 s, dry analytical flow cell < ±3.5 μAU (typical < ±2.5 μAU) at 254 nm, time constant 1 s, dry analytical flow cell Noise, multiple wavelength (VWD-3400RS): < ±10 µAU (typical < ±7 µAU) at 254 nm and 280 nm, 2 s time constant, dry analytical flow cell Drift: < 10-4 AU/h at 254 nm and dry analytical flow cell Linearity: < 5% RSD (typical < 3% RSD) at 2.5 AU caffeine, wavelength: 272 nm based on ASTM Light source: Deuterium lamp, tungsten lamp (Tungsten lamp optional on VWD-3100) Temperature control for both lamps Wavelength range: 190–900 nm The tungsten lamp is recommended for wavelengths > 600 nm Wavelength accuracy: ±1 nm (over detector lifetime) Wavelength repeatability: ±0.1 nm Optical bandwidth: 6 nm at 254 nm I/O interfaces: Four digital inputs, four digital outputs Analog output: Two analog outputs available as an option via DAC plug-in module Software selectable: absorbance, 20-bit resolution, 0–1 V (full range) and 0–10 V (full range with adjustable mAU ranges). Safety feature: Power-up diagnostics of optics, cooling fans, motors, and electronics leak sensor Power consumption: 85–260 V AC, 50/60 Hz, max. 150 W Wide range (automatic voltage selection) Wetted parts: PEEK, quartz glass, stainless steel Dimensions (h × w × d): 16 × 42 × 51 cm (6.3 × 16.5 × 20 in.)

Quote notes

Weight 0,0 kg
Condition

As good as new

Brand

Dionex

Main Menu